User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: DATA TECH 40 DATE: 22-Apr-87 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 29 NUMBER OF LINES: 59 CONFIGURATION: Fluke 5100B Fluke 5205A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD DC VOLTAGE TESTS 1.002 DISP Set the T.I. to ON. 1.003 DISP Allow 15 minutes warm-up time 1.004 DISP Set the T.I. FUNCTION switch to DCV 1.005 DISP Connect the leads to the T.I. HIGH and LOW terminals. 1.006 5100 100 100mV 0.08U 2W 2.001 5100 1 1V 0.0006U 2W 3.001 5100 1 -1V 0.0006U 2W 4.001 5100 10 10V 0.006U 2W 5.001 5100 10 9V 0.006U 2W 6.001 5100 10 8V 0.005U 2W 7.001 5100 10 7V 0.005U 2W 8.001 5100 10 6V 0.005U 2W 9.001 5100 10 5V 0.005U 2W 10.001 5100 10 4V 0.004U 2W 11.001 5100 10 3V 0.004U 2W 12.001 5100 10 2V 0.004U 2W 13.001 5100 10 1V 0.003U 2W 14.001 5100 100 100V 0.06U 2W 15.001 5205 1000 1000V .6U E 2W 16.001 HEAD AC VOLTAGE TESTS 16.002 DISP Set the T.I. FUNCTION switch to ACV. 16.003 5100 1 1V 0.0038U 400H 2W 17.001 5100 10 10V 0.048U 400H 2W 18.001 5100 100 100V 0.48U 400H 2W 19.001 5100 1000 100V 1.3U 400H 2W 20.001 5100 1 1V 0.0038U 5KH 2W 21.001 5100 10 10V 0.048U 5KH 2W 22.001 5100 100 100V 0.48U 5KH 2W 23.001 5100 1000 100V 1.3U 5KH 2W 24.001 HEAD OHMS TESTS 24.002 DISP Set the T.I. FUNCTION switch to K-OHM 24.003 RESF 1 1KZ 0.0013U 2W 25.001 RESF 10 10KZ 0.013U 2W 26.001 RESF 100 100KZ 0.13U 2W 27.001 RESF 1 1MZ 0.0013U 2W 28.001 RESF 10 10MZ 0.013U 2W 29.001 DISP Disconnect all leads and turn the T.I. to OFF. END